    <title locale="en_US">Effect of Abutment Taper on the Fracture Resistance of all-Ceramic Three-unit Bridges</title>
    <FirstPage>159</FirstPage>
    <LastPage>167</LastPage>
    <AuthorList>
      <Author>
        <FirstName></FirstName>
        <LastName>F. Gerami-Panah</LastName>
        <affiliation locale="en_US"></affiliation>
      </Author>
      <Author>
        <FirstName></FirstName>
        <LastName>H. Jalali</LastName>
        <affiliation locale="en_US"></affiliation>
      </Author>
      <Author>
        <FirstName></FirstName>
        <LastName>L. Sedighpour</LastName>
        <affiliation locale="en_US"></affiliation>
      </Author>
    </AuthorList>
    <History>
      <PubDate PubStatus="received">
        <Year>2015</Year>
        <Month>10</Month>
        <Day>03</Day>
      </PubDate>
    </History>
    <abstract locale="en_US">Statement of Problem: The connector area is the weakest zone of an all-ceramic fixed partial denture (FPD), where most catastrophic failures of the prostheses tend to occur.
Purpose: The aim of the present study was to assess the effect of the convergence angle of abutments on the fracture resistance of three-unit fixed partial dentures made of IPS-Empress2.
Materials and Methods: Forty extracted human premolars and molars were used to reproduce twenty, 3-unit fixed partial dentures, for the replacement of second premolars. All teeth were prepared according to the guidelines outlined for all-ceramic crowns and bridges, except for the convergence angles of the abutments. The specimens were randomly divided into two groups of 10, with total occlusal convergence angles of 12&#xB0; and 22&#xB0;. Fixed partial dentures with a uniform thickness of 0.8 mm were fabricated using IPS-Empress2 and were bonded to the corresponding models. Connector dimensions were set to 4 mm height and 4mm width. The radius of curvature at the gingival embrasure was carved to 0.9 mm. All specimens were exposed to 10,000 pre-loading cycles and a load of 40 N at a frequency of 1.3 Hz in a standardized testing machine at a cross head speed of 1mm/min. Student t-test was performed to detect any difference in the mean fracture resistance between the two groups (&#x3B1; = 0.05).
Results:  Mean failure loads (and standard deviations) of the 12&#xB0; and 22&#xB0; groups were 1009.12 N (208.05) and 1182.72 N (144.67), respectively. Statistical analysis revealed a significant difference (P</abstract>