Accuracy of an Electronic Apex Locator for Working Length Determination in Primary Anterior Teeth

Abstract

Objectives: Correct determination of working length is an important step for suc- cess of endodontic therapy. Conventional radiography has limitations in providing the accurate location of apical foramen. For this reason, electronic apex locators (EALs) were developed to shorten the treatment time and decrease the radiation dose. The aim of this study was to evaluate the accuracy of Root ZX EAL for work- ing length determination in primary anterior teeth.

Materials and Methods: In this in-vitro study, 50 extracted primary anterior teeth with root resorption were selected with no obstructed canals or history of previous root canal therapy. Working lengths were measured by direct observation of actual length (AL), radiography and Root ZX EAL. A variation of ±0.5 mm from the AL was considered acceptable. The results were analyzed statistically using paired t test and interclass correlation coefficient (ICC) at 0.05 level of significance.

Results: Considering an acceptable 0.5 mm margin from AL (direct measurement), the accuracy of Root ZX EAL and radiography was found to be 86% and 76%, respectively. Absolute value of error from AL was significantly lower for ZX com- pared to radiography (P˂0.001). Interclass correlation comparing both radiography and Root ZX with AL showed strong correlations.

Conclusion: Root ZX EAL can be used as a reliable tool for obtaining root canal length in primary anterior teeth with resorption.

Files
IssueVol 12, No 4 (2015) QRcode
SectionOriginal Article
Keywords
Deciduous tooth Radiography Tooth apex

Rights and permissions
Creative Commons License This work is licensed under a Creative Commons Attribution-NonCommercial 4.0 International License.
How to Cite
1.
Bahrololoomi Z, Soleymani A, Modaresi J, Imanian M, Lotfian M. Accuracy of an Electronic Apex Locator for Working Length Determination in Primary Anterior Teeth. Front Dent. 2015;12(4):243-248.