Microleakage Evaluation at Implant-Abutment Interface Using Radiotracer Technique
Abstract
Objectives: Microbial leakage through the implant-abutment (I-A) interface results in bacterial colonization in two-piece implants. The aim of this study was to compare microleakage rates in three types of Replace abutments namely Snappy, GoldAdapt, and customized ceramic using radiotracing.
Materials and Methods: Three groups, one for each abutment type, of five implants and one positive and one negative control were considered (a total of 17 regular body implants). A torque of 35 N/cm was applied to the abutments. The samples were immersed in thallium 201 radioisotope solution for 24 hours to let the radiotracers leak through the I-A interface. Then, gamma photons received from the radiotracers were counted using a gamma counter device. In the next phase, cyclic fatigue loading process was applied followed by the same steps of immersion in the radioactive solution and photon counting.
Results: Rate of microleakage significantly increased (P≤0.05) in all three types of abutments (i.e. Snappy, GoldAdapt, and ceramic) after cyclic loading. No statistically significant differences were observed between abutment types after cyclic loading.
Conclusions: Microleakage significantly increases after cyclic loading in all three Replace abutments (GoldAdapt, Snappy, ceramic). Lowest microleakage before and after cyclic loading was observed in GoldAdapt followed by Snappy and ceramic.
- Salihoglu U, Boynuegri D, Engin D, Duman AN, Gokalp P, Balos K. Bacterial adhesion and colonization differences between zirconium oxide and titanium alloys: an in vivo human study. Int J Oral Maxillofac Implants. 2011 Jan;26(1):101-7.
- Harder S, Dimaczek B, Açil Y, Terheyden H, Freitag-Wolf S, Kern M. Molecular leakage at implant-abutment connection in vitro investigation of tightness of internal conical implant-abutment connections against endotoxin penetration. Clin Oral Invest. 2010 Aug;14(4):427-32.
- da Silva-Neto JP, Nobilo MA, Penatti MP, Simamoto PC Jr, Neves FD. Influence of methodologic aspects on the results of implant- abutment interface microleakage tests: a critical review of in vitro studies. Int J Oral Maxillofac Implants. 2012 Jul-Aug;27(4):793-800.
- Broggini N, McManus LM, Hermann JS, Medina RU, Oates TW, Schenk RK, et al. Persistent acute inflammation at the implant-abutment interface. J Dent Res. 2003 Mar;82(3):232-7.
- Steinebrunner L, Wolfart S, Bössmann K, Kern M. In vitro evaluation of bacterial leakage along the implant- abutment interface of different implant systems. Int J Oral Maxillofac Implants. 2005 Nov- Dec;20(6):875-81.
- do Nascimento C, Barbosa RE, Issa JP, Watanabe E, Ito IY, Albuquerque RF Jr. Bacterial leakage along the implant-abutment interface of premachined or cast components. Int J Oral Maxillofac Surg. 2008 Feb;37(2):177-80.
- Silva-Neto JP, Prudente MS, Carneiro Tde A, Nóbilo MA, Penatti MP, Neves FD. Micro-leakage at the implant-abutment interface with different tightening torques in vitro. J Appl Oral Sci. 2012 Sep- Oct;20(5):581-7.
- Aníl N, Hekimoglu C, Büyükbas N, Ercan MT. Microleakage study of various soft denture liners by autoradiography: effect of accelerated aging. J Prosthet Dent. 2000 Oct;84(4):394-9.
- Kano SC, Binon PP, Curtis DA. A classification system to measure the implant-abutment microgap. Int J Oral Maxillofac Implants. 2007 Nov-Dec;22 (6):879-85.
- Sarac D, Sarac YS, Basoglu T, Yapici O, Yuzbasioglu E. The evaluation of microleakage and bond strength of a silicone- based resilient liner following denture base surface pretreatment. J Prosthet Dent. 2006 Feb;95(2):143-51.
- Albert FE, El-Mowafy OM. Marginal adaptation and microleakage of procera AllCeram crowns with four cements. Int J Prosthodont. 2004 Sep-Oct;17(5):529-35.
- McCurdy CR Jr, Swartz ML, Phillips RW, Rhodes BF. A comparison of in vivo and in vitro microleakage of dental restorations. J Am Dent Assoc. 1974 Mar;88(3):592-602.
- Teixeira W, Ribeiro RF, Sato S, Pedrazzi V. Microleakage into and from two-stage implants: an in vitro comparative study. Int J Oral Maxillofac Implants. 2011 Jan-Feb;26(1):56-62.
- Scarano A, Assenza B, Piattelli M, Iezzi G, Leghissa GC, Quaranta A, et al. A 16-year study of the microgap between 272 human titanium implants and their abutments. J Oral Implantol. 2005 Dec;31 (6):269-75.
- Charlton DG, Moore BK. In vitro evaluation of two microleakage detection tests. J Dent. 1992 Feb;20 (1):55-8.
- Dias EC, Bisognin ED, Harari ND, Machado SJ, da Silva CP, Soares GD, et al. Evaluation of implant- abutment microgap and bacterial leakage in five external-hex implant systems: an in vitro study. Int J Oral Maxillofac Implants. 2012 Mar-Apr;27(2):346-Am Dent Assoc. 1964 Oct;69:416-22.
- Andersson B, Odman P, Lindvall AM, Brånemark PI. Cemented single crowns from a prospective study on CeraOne. Int J Prosthodont.1998 May-Jun;11(3):212-8.
- Rekow D, Thompson VP. Engineering long term clinical success of advanced ceramic prostheses. Journal of Materials Science. J Mater Sci Mater Med.2007 Jan;18(1):47-56.
- Sandler MP. Diagnostic nuclear medicine. 4th ed. Philadelphia, Lippincott Williams & Wilkins,2003:9-15.
- Coelho PG, Sudack P, Suzuki M, Kurtz KS, Romanos GE, Silva NR. In vitro evaluation of the implant abutment connection sealing capability of different implant systems. J Oral Rehabil. 2008
Dec;35(12):917-24.
- Jansen VK, Conrads G, Richter EJ. Microbial leakage and marginal fit of the implant-abutment interface. Int J Oral Maxillofac Implants. 1997 Jul- Aug;12(4):527-40.
- Hjerppe J, Lassila LV, Rakkolainen T, Narhi T, Vallittu PK. Load-bearing capacity of custom-made versus prefabricated commercially available zirconia abutments. Int J Oral Maxillofac Implants. 2011 Jan- Feb;26(1):132-8.
- Going RE. Cavity liners and dentinal treatment. J Am Dent Assoc. 1964 Oct;69:416-22.
- Andersson B, Odman P, Lindvall AM, Brånemark PI. Cemented single crowns from a prospective study on CeraOne. Int J Prosthodont.1998 May-Jun;11(3):212-8.
- Rekow D, Thompson VP. Engineering long term clinical success of advanced ceramic prostheses. Journal of Materials Science. J Mater Sci Mater Med.2007 Jan;18(1):47-56.
- Sandler MP. Diagnostic nuclear medicine. 4th ed. Philadelphia, Lippincott Williams & Wilkins,2003:9-15.
Files | ||
Issue | Vol 13, No 3 (2016) | |
Section | Original Article | |
Keywords | ||
Microleakage Dental Implant Implant-Abutment Interface Connections Radiotracer |
Rights and permissions | |
This work is licensed under a Creative Commons Attribution-NonCommercial 4.0 International License. |