Comparison of Implant Stability Using Resonance Frequency Analysis: Osteotome versus Conventional Drilling
Abstract
Objectives: Implant primary stability is one of the important factors in achieving implant success. The osteotome technique may improve primary stability in patients with poor bone quality. The aim of this study was to compare implant stability using two different techniques namely osteotome versus conventional drilling in the posterior maxilla.
Materials and Methods: In this controlled randomized clinical trial, 54 dental implants were placed in 32 patients; 29 implants were placed in the osteotome group and 25 in the conventional drilling group. Implant stability was assessed at four time intervals namely at baseline, one, two and three months after implant placement using resonance frequency analysis (RFA).
Results: Primary stability based on implant stability quotient (ISQ) units was 71.4±7 for the osteotome group and 67.4±10 for the control group. There was no statistically significant difference between the two groups in implant stability at any of the measurement times. In each group, changes in implant stability from baseline to one month and also from two months to three months post-operatively were not significant but from one month to two months after implant placement, implant stability showed a significant increase in both groups.
Conclusion: The results of this study revealed that in both techniques, good implant stability was achieved and osteotome technique did not have any advantage compared to conventional drilling in this regard.
Files | ||
Issue | Vol 12, No 9 (2015) | |
Section | Original Article | |
Keywords | ||
Dental Implants Bone Maxilla Osteotomy |